Edit METHOD SETUP
Edit
METHOD SETUP
No. of Decimals
1
2
Printout Format
RESULT
ALL
Warning Beep
YES
NO
Result Beep
YES
NO
Return to the first Edit display by pressing the up arrow key or leave the Edit
mode by pressing the Mode, Cal, Sample, pH/mV, or GLP key.
Using:
Select METHOD SETUP in the Edit display
Press the down arrow key to continue.
Using:
Select the number of decimals to be displayed:
pH: 1 or 2; mV: 0 or 1.
Using:
Select the printout format you require for sample
measurements.
If you choose RESULT: the sample number,
pH/mV result, temperature and stabilisation time
are printed out. For manual measurements the
pH/mV result and temperature are printed out.
If you choose ALL: calibration data, i.e. zero pH
and sensitivity, will be included in the printout;
refer to page 44.
Using:
Select YES, if you want a warning beep (three
short tones) to accompany an alarm message.
Select NO, If no beep is required.
Using:
Select YES, if you want a result beep
(one long tone) to sound when a result
has been obtained.
Select NO, if no beep is required.
The EDIT mode
41