3.
The PHM220 will now measure and await electrode stability. The
measurement can be followed by means of the stability indicator.
Once stable, or after a preset time period, the mV value will be locked
on the display.
4.
As soon as the mV value has been accepted, the following display will
appear:
The result is automatically stored in the GLP result table.
Once the result has been accepted, it will be automatically printed out if
a printer is connected. In addition to the mV value, the printout will
show the continuously incremented sample number and the measured
or adjusted sample temperature.
5.
A new sample measurement is started by pressing the
the pH/mV key to change to the live display, if desired.
In order to carry out a mV measurement AT INTERVALS, it is necessary to
select AT INTERVALS in Edit SAMPLE and enter the number of seconds or
minutes desired between each measurement and result printout. A mV
measurement at intervals takes place in the same way as a pH measure-
ment at intervals; refer to page 24.
26
Chapter 2
STA
STA
20s
RESULT
No.18
Measurements AT INTERVALS
134.2mV
22.0˚C
134.2mV
22.0˚C
key. Press