Platinum Electrode Operating Instructions The M241Pt Platinum Electrode is ideal for a wide range of ap- plications. It can be used together with a standard Ag/AgCl ref- erence electrode or, depending on the application, with a refer- ence electrode which has an alternative liquid junction or refer- ence system.
Cleaning and maintenance Platinum plate The sensing part of the electrode can be polished with finishing abrasive bands, part no. B90X550. While polishing, take great care not to bend the platinum plate. In order to have reproducible and accurate results with the electrode, a few physical and electrochemical treatments are required.
Accessories Description Part No. BA03 Finishing Abrasive Bands B90X550 RENOVO•N Normal Cleaning Solution, 250 ml S16M001 Specifications Temperature range: 0 to 100°C Sensing element: M241Pt : platinum plate 5 x 5 mm Connection: Banana plug...
Page 6
Electrode de Platine Mode d'Emploi L'électrode de platine M241Pt convient à de nombreuses applica- tions. Elle peut être utilisée avec une électrode de référence Ag/ AgCl classique ou, en fonction de l'application, avec une électrode de référence possédant une jonction liquide ou un système de référence différent.
Page 7
Entretien Plaque de platine La surface active de l'électrode peut être préparée par polissage avec des bandes abrasives de finition, réf no. B90X550. Faîtes très attention lors du polissage à ne pas plier la plaque de platine : risque de casser la soudure verre-platine. Pour obtenir une électrode bien définie et reproductible, il est nécessaire de lui faire subir divers traitements physiques et électrochimiques qui sont en fonction des manipulations anté-...
Page 8
Accessoires BAO3 Bandes Abrasives de Finition B90X550 RENOVO•N Solution pour Nettoyage Normal, 250 ml S16M001 Spécifications Gamme de température : 0 à 100 °C Partie active : M241Pt : plaque de platine (5 x 5 mm) Connexion : Fiche banane...