The EDIT Mode

A pH method or a mV method can be edited to fit your specific tasks.
The Edit mode has four options: CALIBRATION, SAMPLE, METHOD
SETUP and CONFIGURATION. Select the options in the Edit display
using the left or right arrow key.
An overview of the displays with a step-by-step explanation of the
editing procedure for each option starts on page 33.
Entering the Edit Mode
Press the Mode key. Use the left or right arrow key to select the
method to be edited: pH or mV.
Enter the Edit mode by pressing the Mode key for 4 seconds.
Two options are available in the Reset Method display which now
appears:
Reset pH method
YES
Reset pH confirm
YES
Mode
Mode
22MAR95 11:15:06
If NO: The present values stored in
the method will remain the same.
NO
When the down arrow key is
pressed, the first edit option will be
displayed. From here you can edit the
whole method or only parts of it to suit
your particular requirements.
If YES: The method calibration data
will be reset to default values. A list of
NO
these values is shown on page 61.
Therefore, as a precaution, a second
display appears when you press the
down arrow key to allow you to
confirm your selection. Now press the
down arrow key and the first edit
display will appear.
pH
The EDIT mode