Operating Manual
Chapter 5

Calibration

Calibrating the Orion Multigas Detector

Each Orion Multigas Detector is equipped with an autocalibration
feature to make unit calibration as easy as possible.
The Autocalibration sequence resets instrument zeroes and
adjusts sensor calibration for known concentrations of
calibration gases.
Table 5-1. Autocalibration and Required Calibration Cylinders
EXPECTED GAS
SENSORS
CONCENTRATION
Combustible
58% LEL
Oxygen
15%
Carbon Monoxide
300 ppm
Hydrogen Sulfide
10 ppm
To Calibrate the Orion Multigas Detector
(FIGURE 5-1):
1. Turn ON the instrument and verify that battery is
sufficiently charged.
2. Push and hold the CHANGE/RESET button until the
ZERO flag flashes in the upper portion of the display
(FIGURE 5-2).
• Indicates instrument is in the Calibration mode.
3. Push the ON-OFF/PAGE button to zero the instrument.
• You must be in fresh air to perform the zero.
• The ZERO flag stops flashing and remains ON.
NOTE: To skip the zero procedure and move directly to the
calibration span procedure, push the
CHANGE/RESET button. If no button is pushed for
10 seconds, the instrument returns to the
Measure mode.
• Once the zeros are set, CAL flag flashes (FIGURE 5-3).
Chapter 5, Calibration
FOUR
GAS CYLINDER
GAS CYLINDER
(P/N 804770, 711058)
(P/N 10010162)
THREE
5-1