MSA Orion Mode D'emploi page 77

Détecteur multi-gaz
Masquer les pouces Voir aussi pour Orion:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Operating Manual
Table 7-6. CARBON MONOXIDE (appropriate models only) -
Typical Performance Specifications
RANGE
999 ppm CO
RESOLUTION
1 ppm CO
REPRODUCIBILITY +2 ppm CO or 10% of reading, whichever is greater
90% of final reading in 30 seconds (normal temperature
RESPONSE TIME
range)
Table 7-7. CARBON MONOXIDE - Cross Reference Factors for Orion
Calibration Using Calibration Cylinder (P/N 478191) or (P/N 804770)
NOTE: Data is presented as the indicated output in ppm,
which would result from the application of 100 ppm of the test gas.
TEST GAS (100 PPM)
Carbon Monoxide (CO)
Hydrogen Sulfide (H
S)
2
Sulfur Dioxide (SO
)
2
Nitrogen Dioxide (NO
)
2
Nitric Oxide (NO)
Chlorine (Cl
)
2
Ammonia (NH
)
3
Hydrogen Chloride (HCl)
Ethylene (C
H
)
2
4
Hydrogen Cyanide (HCN)
Methane (CH
)
4
Ethanol (EtOH)
Hydrogen (H
)
2
Chapter 7, Performance Specifications
*See TABLE 7-2, NOTE 1.
EQUIVALENT PPM
100 +9
1 +6
0 +1
2 +6
70 +10
1 +8
2 +4
3 +2
80 +9
0 +1
0 +0
4 +5
70 +26
7-5

Publicité

Table des Matières
loading

Table des Matières