Cleaning and Maintenance

Procedure
Carry out weighing .
t
Press the [Start GLP] button .
t
If the "Batch ID" function is activated: Enter a batch
t
ID .
If the "Sample ID" function is activated: Enter a
t
sample ID .
The GLP header is printed with the device ID, the
y
batch ID, the sample ID, and the current weight
value .
To exit the GLP printout: Press the [Exit GLP] button .
t
The GLP footer is printed .
y
9
Cleaning and
Maintenance
9 . 1
Detaching the Analytical Draft
Shield
Material:
1 soft support base
Requirements
The device is disconnected from the power supply .
Procedure
Place the side panels and the upper panel on a soft
t
surface .
If an analytical balance is to be cleaned: Remove the
t
weighing pan and the shield disk .
If a precision balance is to be cleaned: Remove the
t
weighing pan .
42
Entris
II Advanced Line Operating Instructions
®
To detach the handles
t
on the side panels and
the upper panel:
Loosen the screws on
the inner sides of the
handles by turning .
CAUTION Glass
t
breakage due to
incorrect handling!
Slide the side panels
t
and upper panel all
the way back and out
of the guide rails .
9 .2

Cleaning the Device

Procedure
Disconnect the device from the power supply .
t
Only use suitable cleaning agents and cleaning
t
procedures and observe the product information for
the cleaning agent used (for cleaning agent, see
Chapter "14 .7 Cleaning Agent", page 49) .
Remove dust and powdery sample residue with a
t
brush or hand-held vacuum cleaner .
Wipe the device and the associated components
t
with a slightly damp cloth . Use a mild soapy solution
or a suitable cleaning agent for more severe
contamination .
9 .3
Installing the Analytical Draft
Shield
Procedure
If an analytical balance is used: Put the shield disk
t
and the weighing pan into place .
If a precision balance is used: Position the weighing
t
pan .
Connect the device to the power supply (see
t
Chapter 6.2, page 36).
9 .4

Maintenance Schedule

Interval
Component
Regularly;
Device
depending on
the operating
conditions
9 .5

Software Update

For a software update, contact Sartorius Service .
Slowly slide the side
t
panels and the upper
panel into the frame
rails .
Hold the handles on
t
the side panels or the
upper panel from the
outside and fasten to
the inner side with the
screws .
Action
Perform a
functional test of
the device .
Contact Sartorius
Service for this .