8.12 Printing Weighing Result with ID Marking
The device, the sample and a batch can be assigned an ID
number . The ID numbers are exported during GLP-compliant
printing .
Requirements
− The identity number is specified (see Chapter "7 .3 .20
"INPUT" Menu", page 27) .
− The printing of the lot ID line in the GLP printout is
activated in the menu (see Chapter "7 .3 .20 "INPUT" Menu",
page 27) .
− The "SPL . ID" output is activated in the menu .
− The GLP-compliant printout is activated (see Chapter "7 .3 .7
"DATA .OUT ."/"PRNT .PAR ." Menu", page 23) .
Procedure
Start the printout . To do so, press the [Print] button .
t
The GLP header is printed with the ID marking of the
y
LOT ID set in the menu and the current weight value .
The [GLP] button appears in the operating display .
y
Confirm the [PRINT] button .
t
The SPL . ID and the current weight value are exported .
y
Confirm the [PRINT] button .
t
The SPL . ID and the current weight value are exported .
y
To exit the GLP printout: Press the [GLP] button .
t
The GLP footer is printed .
y
9
Cleaning and
Maintenance
9.1
Detaching the Draft Shield
9.1.1
Detaching the Analytical Draft Shield, Weighing
Pan, and Associated Components
Material:
1 soft support base
Requirements
− The device is switched off .
− The device is disconnected from the power supply .
Procedure
Slide the side panels and upper panel all the way back and
t
out of the guide rails .
Place the side panels and the upper panel on a soft surface .
t
If this relates to a device with a shield disk: Remove the
t
weighing pan, the pan support, and the shield disk .
If this relates to a device without a shield disk: Remove the
t
weighing pan and the pan support .
9.1.2
Dismantle the Round Draft Shield, Weighing Pan,
and Associated Components
Material:
1 soft support base
Requirements
− The device is switched off .
− The device is disconnected from the power supply .
Procedure
Entris
®
Cleaning and Maintenance
Loosen the screws on the
t
inner side of the side panels
by turning . This allows the
handles on the side panels
and the upper panel to be
removed .
CAUTION Glass breakage
t
due to incorrect handling!
Remove the cover and the
t
round glass ring and place
on a soft surface .
Remove the weighing pan,
t
pan support, and the base of
the draft shield .
BCE Operating Instructions
31