Connection
It is essential to fit the power cable with a plug that is suitable for the above power.
We recommend using the following kind of plug, which is especially suited to frequent disconnection and
reconnection operations:
We recommend positioning the wall socket between 1 and 1.20 meterss from the ground and fitting it with a
nearby cut-off device, in order to carry out connection and disconnection operations with the power off.
Water
Supply
•
For more convenient use, include an easy to access tap close to the MEDIAWEL, at a height of
approximately 1 meter and slightly offset at the side of the device as illustrated below.
•
Cut the 4 meter hose to the desired lengths for supply and outlet hoses
•
Fit the in-line filter to the water supply hose
•
Connect the supply and outlet hoses as set out below
Marking
Name
1
In-line water circuit filter (recommended option)
2 and 3
Inlet and outlet water tubing
Draining outlet
3
Part No.
MEDW3008
MEDW4012
69
2
Water supply
Water filter
1