Alliance Bio Expertise MEDIAWEL 30 Manuel D'utilisation page 110

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CYCLE
PROGRAM
STIRRER
WATER
PARAMETERS
16:46:35
Message :
Here below are listed the messages that can be displayed:
Message
Step Start Alarm
Alarm 'Close Vessel Lid'
Alarm 'Close Safety Covers'
Heating error \n\n Abnormal temperature
variation
Cooling error \n\n Abnormal temperature
variation
Temperature warning \n\n temperature is
out of tolerance
Restart cycle after power on
## : Name of Program
Warning message area
Action
Follow the instruction of the message that have been
created for this message. Example:"Add blood"
The lid is not properly closed. Release the pressure from
the vessel by using the safety while the temperature is
below 80°C. Then open the lid and close it again to get
blue light coming one. If not
One of the safety lids (door trap or lid to access to the
dispensing port) is not closed or not properly closed-->
Close the lids
Time to reach the sterilization set point is too long or the
temperature is not stable --> Make sure that the stirrer is
running properly. If not, check the bearing state.
Time to reach the dispensing set point is too long or the
temperature is not stable --> Make sure that the water
tap is open.
The temperature was not in +/-1°C range during the
sterilization plateau- Make sure that temperature is
within the tolerance of your application before validating
the cycle.
Just for information - Check the data records at the end
of the cycle before validating the cycle.
110
Current
stepXXX
→Call service.