End of the cycle and dispensing
Sequence
of
operations
1
-
-
2
-
3
-
4
-
5
-
-
6
-
The end of the cycle is indicated by the lights changing from blue to
green
This means that the dispensing temperature has been reached and at
its step → dispensing can be undertaken
Open the safety cover, by operating the small lever on the right, and
lift it up
Pull the ring on the safety valve in order to ensure that any possible
pressure is released
Prepare the decanting tubing and place it on top of the lid but leave
it in its protective packaging before removing the decanting cap
Heat the adding port with a blow torch before inserting the tip of the
decanting tubing
Do not heat the decanting tip to not damage the seal
Connect the hose to the pump and run dispensing immediately
Operations
103
Symbols