Alliance Bio Expertise MEDIAWEL 30 Manuel D'utilisation page 111

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

'Media' temperature probe error \n\n
Abnormal temperature measurement
'Heating' temperature probe error \n\n
Abnormal temperature measurement
Technical Error \n\n Heating Elements
Technical Error \n\n Stirrer motor
Technical Error \n\n Pressure release
solenoid
Technical Error \n\n Cooling cooling
Technical Error \n\n Locker magnet
Technical Error \n\n Other\
Communication Initialisation failed
Communication failed. Please switch off
and contact the after-sales service
The vessel temperature probe is defective --> Call service
The heating element temperature probe is defective -->
Call service
Defective heating element --> Call service
Defective stirrer --> Call service
Defective pressure release valve --> Call service
Defective cooling valve --> Call service
Defective lid locking solenoid --> Call service
Call service
Switch off and on again the machine - Call service if it
does not solve the problem
Switch off and on again the machine - Call service if it
does not solve the problem