Connection; Water; Power Supply; Initial Commissioning - Alliance Bio Expertise MEDIAWEL 10 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Connection

It is essential to fit the power cable with a plug that is suitable for the above power.

Water

Inlet
For more convenient use, include an easy to access tap close to the MEDIAWEL, at a height of
approximately 1 meter and slightly offset at the side of the device as illustrated below
Cut the 4 meter hose to the desired lengths for supply and outlet hoses
Fit the in-line filter (ref: MEDW3008) to the water supply hose
Connect the supply and outlet hoses as set out below
Repère
Désignation
1
In-line water circuit filter (recommended option)
2 et 3
Inlet and outlet water tubing
Outlet
The cooling water drainage system must be designed to withstand a temperature of 130°C. It may be made
from copper or PVC-C or any other material that is suitable for this use.

Power supply

The adapted plug must be used adapted to the above specifications :
Phase
-
Neutral
-
Earth
-
The wall power supply must be between 1 and 1,2m height and equipped with a shutdown system.

Initial commissioning

Switching on

Prior to switching the device on, you must ensure that the mains voltage between phase and neutral
complies with 230V +/- 10%. Plug the machine using the plug into the wall power supply and switch on the
front switch.
1,
wire is identifyed by
wire is identifyed by
wire is yellow and green
Water outlet
2
3
printed on one of the 2 black cables
2,
printed on the second black cable
54
1
Water inlet
In-line water supply
(ref : MEDW3008)
Référence
MEDW3008
MEDW4012

Publicité

Table des Matières

Dépannage

loading

Table des Matières