Insufficient Pressure Support Alarm; Flow Blocked Alarm - ResMed AutoSet CS 2 Manuel Utilisateur

Masquer les pouces Voir aussi pour AutoSet CS 2:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

eng.book Page 33 Friday, November 3, 2006 2:43 PM
W
ILL STOP WHEN
the mask leak has been rectified.
The High Leak alarm can be turned off in the Leak Alert option in the Settings menu.
N
OTE
I
NSUFFICIENT PRESSURE SUPPORT ALARM
T
RIGGERED BY
air pressure at the mask has failed to reach an expected level for 3 breaths
the Air Circuit Learn function (page 26) has not been run after adding a new
component (eg new mask, humidifier) to the system.
I
I
F AN
NSUFFICIENT PRESSURE SUPPORT ALARM SOUNDS
check that the pressure sensor tubing is not kinked
check that the pressure sensor tubing is properly connected
run LEARN CIRCUIT.
W
ILL STOP WHEN
the pressure sensor tubing is unobstructed and properly connected to the flow
generator and mask
treatment is stopped (Start/Stop key or S
the A
S
UTO
ET
F
LOW BLOCKED ALARM
T
RIGGERED BY
blockage in air circuit
I
F
F A
LOW BLOCKED ALARM SOUNDS
1. The treatment will stop.
2. Turn power off.
3. Check whether there is a blockage in the air circuit.
4. Remove blockage.
5. Turn power back on.
6. If the Flow blocked alarm activates repeatedly, discontinue use and return to
ResMed for servicing. If the alarm does not recur, then continue to use as normal.
W
ILL STOP WHEN
blockage is removed.
:
:
:
CS2 is turned off.
:
:
:
S
)
MART
TOP
:
T
A
HE
LARMS
33