2
Connect air tubing
Connect the air tube firmly onto the air outlet at the front of the unit.
Figure 4: Front view of AutoSet CS2
To attach the pressure sensor tube to the unit, screw the Luer connector onto the
socket on the side of the unit.
Figure 5: Twisting the Luer lock connectors together
WARNING
!
Use care when handling the air tubing. Be particularly careful when attaching the
Luer connector to the A
tube, it may affect the protection offered by the mask pressure alarm system.
12
Air tubing connector
Pressure sensor tube
connector
S
CS2. If you kink or excessively twist the sensor
UTO
ET