Safety - Haag-Streit Goldmann AT 900 Mode D'emploi

Tonomètre à aplanation

Publicité



Safety

Ambient conditions
• Transportation Temperature
Air pressure
Relative humidity 10%
• Storage
Temperature
Air pressure
Relative humidity 10%
• Working
Temperature
Air pressure
Relative humidity 30%
Installation, dispatch
• To avoid condensation, allow the instrument to adjust to room
temperature for several hours before unpacking.
• The Applanation Tonometer should be verified for damage
after unpacking.
• Installation is to be carried out according to Appendix A and
functional verifications according to Appendix B.
• Check: do the connection pieces fit (tonometer on slit lamp,
measuring prism)?
• Defective equipment should always be returned in an ap-
propriate packaging.
Operation, surrounding
• Only qualified and trained personnel should operate the equip-
ment, the training is at the owner's responsibility.
• Use only Haag-Streit accessories.
• Avoid humidity.
Cleaning of housing
• Use a slightly moistened cloth.
• Soap can be employed.
• No liquids or corrosive agents.
• Avoid direct spraying on the equipment.
Warranty, product liability
• The instrument should be operated in accordance with the
chapter 'Safety'. Incorrect operation can damage the instru-
ment. Thus no warranty claims can be accepted.
• In particular, it is forbidden to loosen any screws; this may
reduce the accuracy of the measurement.
• Continued use of an instrument that has been damaged by
incorrect operation can lead to personal injury. The manufac-
turer cannot accept liability in this case.
• Repairs and alterations on this equipment should only be car-
ried out by Haag-Streit. Contact your Haag-Streit distributor.
-40 °C
to +70 °C
500 hPa to 1060 hPa
to 95%
-10 °C
to +55 °C
700 hPa to 1060 hPa
to 95%
+10 °C to +35 °C
800 hPa to 1060 hPa
to 75%
1500.7006000.04091
© Haag-Streit AG, CH-3098 Koeniz, Switzerland
5