Maintain the tonometer
Haag-Streit AG
Gartenstadtstrasse 10
CH-3098 Koeniz, Switzerland
Phone ++ 41 31 978 01 11
Fax
++ 41 31 978 02 82
eMail
info@haag-streit.ch
internet www.haag-streit.com
© Haag-Streit AG, CH-3098 Koeniz, Switzerland
How to disinfect?
carefully measuring prism from holder
Pantasept
®
0 minutes
3 % aqueous solution
• Clean tonometer only with a dampened cloth
• Use soap only
Remove
Clean
30 - 60 seconds
running,
cold water
Disinfect
(Haag-Streit)
or
minutes
%
Rinse
min. 0 - max. 60 minutes
running,
cold drinking water
Dry
With a one-way tissue
clean,
soft
Store
Place into container
clean,
dry
*1500700600004091*
1500.7006000.04091
06.07 - 2
See also: Appendix B.3
Recommendation: Use the
Haag-Streit 'DESINSET'
2.4 Accessories (page 13)
• sensitive feeler arm!
Allowed
• mild soap
The use of the fol-
lowing products /
procedures for di-
sinfection will de-
grade the tip and
cause it to fail:
• Alcohol
• Acetone
• UV radiation
• Sterilization
• Immersing in
fluid for more
than 1 hour
• Temperature
above 60 °C
• Do not immerse
tonometer in solution!
• No corrosive agents
• No direct spraying