Pin 6 : Common for 8 opto-coupler outputs
Pin 5 : Output + 5V / 100 mA , non-regulated (protected)
:
Pin 4
Output + 9V / 200 mA , non-regulated (protected)
Pins 2 & 3 : Input for external contact
Use only one tension of 5V on these inputs.
Pins 2/3
Pin 5 (5V)
Example of connection for Switch 2
80
E
Max voltage = 30V and the
max current is 60 mA per
output.
The opto-coupler outputs
must be supplies
externally with negative
voltage to common
emitters (pin 6)
The protection diode is
necessary in case of
inductive charge (electro-
valve, relay, solenoid...)