To save the current configuration temporarily in the D200S unit, use the menu
ransfer (to D200S). The configuration will be
T
D
200S unit.
T
o save the current configuration definitively in the D200S unit, use the Transfer +
S
ave (to D200S). The configuration will be saved definitively in the D200S unit.
4. The configuration of channel 3 will
be as follows :
Setting
Upper Tol. = 0.005
Lower Tol. = 0
Nom
inal Valu
Preset Value = 0
Measuring direction and mode
Increase / positive (+)
Ext. / external measurement
Mo
de
Delta (max-min mode)
Math function
Factor A = 1
Channel A =1
+
Factor B = 1
Channel B =2
5. Active the Delta m
the icon : MinM
6. Turn the part (360°)
7. The display will indicate :
Chan
nel 1 : direct value of p
Chan
nel 2 : direct value of probe B
Chan
nel 3 : delta value of probes
8. Re-initializa
Clear
icon.
saved until the extinction of the
100
E
e =
0
ode by clicking
ax
robe A
A+B
tion is made by the