Foreword
Maintenance and inspection
Introduction
Maintenance – Annually
1
Foreword
Description
Component
Inspection type
1.1
Introduction
Support bracket
Visual inspection of welds
The Casalift 180 is a lightweight and versatile lift device for use in patient and
No damage
geriatric care situations. Use of the lift is facilitated considerably by its spreading
Connection to boom
travel leg design. After use, or for transport by car, it can be dismantled.
Chassis
Visual inspection of welds
This manual was produced based upon the information available on the design
No damage
and operation of the Casalift which was available at time of printing. We reserve
Castors undamaged
the right to make changes due to technical advances.
Function of the brakes
Should you have any questions concerning the Casalift, please contact your spe-
cialist Days Healthcare dealer (see dealer's stamp on the back of the envelope).
Stress test 180.0 kg (Casalift 180E)
Perform metrological control (see page 34)
1.2
Intended use
The Casalift 180 should be use exclusively for the purpose of lifting and moving
11.2.4 Metrological control
persons in need of care. It is only designed to move patients over short
In some countries of the European Union, the following list of measurements are
distances.
prescribed by law (in Germany according to BGV A3).
Improper use will lead to the exclusion of the manufacturer's liability.
It is the responsibility of the owner of the Casalift to determine the situation in
The Casalift may only be used by persons who have received instruction from the
their particular country.
owner, and under consideration of the condition of the patient being transport-
The following measurements may only be made by trained and experienced per-
ed. In borderline cases, or if in doubt, consult the attending doctor.
sonnel using appropriate measuring instruments. Any defects which are estab-
lished must be notified immediately to the appropriate parties and rectified.
Restriction on use!
Patients who are heavier than 180 kg must not be lifted with the Casalift.
Description
1.3
Information on the operating manual
The following measurements must be made acc. to DIN VDE
0751:
1.3.1
Symbols used in the operating manual
Alternative leakage currents
The following symbols are used in the operating manual to identify particular
hazards in using the product or information to simplify usage.
Information on measuring the alternative leakage currents:
Caution!
Typical value:
< 20 µA
This symbol identifies safety instructions which highlight risks in using the prod-
uct.
Inspection point:
Short-circuited mains plug (L+N) to bare, metallic points on
the frame of the charging station (screws, etc.)
Caution!
This symbol identifies safety instructions which highlight risks of crushing and
trapping when using the product.
GB – 20
GB – 34
36
Evaluation
OK
Defective
Evaluation
OK
Defective
Casalift 140E / 180E – 15.10.2010
Casalift 140E / 180E – 15.10.2010
Casalift 180 - SEPTEMBER 2017
Troubleshooting
Safety
Maintenance – Annually
Safety instructions
12
Troubleshooting
Do not open any electronic components, as this will void the warranty and
Note
the manufacturer's liability for the product.
Under this symbol, you can find helpful information on using the product.
Under normal circumstances, the lifting motor can still travel independently
to the bottom position, even when the battery is almost flat.
1.3.2
Information on the operating manual
If you are unable to rectify the fault, please contact your specialist Days Healthcare
Carefully read through the operating manual before assembling the Casalift and
dealer (see dealer's stamp on the back of the envelope).
bring it into service!
Make sure that:
Fault
Possible cause
all persons who are to use the Casalift have read the operating manual.
Boom does not
EMERGENCY STOP but-
move, despite a but-
all persons charged with setting up or maintaining the Casalift always have
ton has been pressed.
ton on the operating
access to the operating manual.
Connectors are not
controls having been
plugged into the control
Damage caused through a failure to follow the information in the operating
pressed.
module properly.
manual is excluded from any and all warranty claims.
No audible or visual
Battery flat.
The illustrations which appear in this operating manual are only examples of
charge reminder sig-
Battery defective.
how to use Days Healthcare patient lifts, and may differ from your actual model.
nal can be detected.
Lifting motor defective.
2
Safety
Control module defec-
tive.
Lifting or spreading
Operating control defec-
2.1
Safety instructions
motor does not re-
tive
spond
2.1.1
Safety instructions – General
Risk of accident!
Only use the Casalift for its intended use. The Casalift is not a piece of gym-
nastics or playground equipment.
Lifting motor defective.
Only use the Casalift if it is in a technically sound condition.
Spreading motor defec-
Only use slings, sling seats or reclined slings which are in perfect condition. Replace
tive
sling systems which are damaged or heavily worn immediately!
Control module defec-
All faults which have a bearing on the function and safety of the Casalift
tive.
must be rectified without delay by a specialist dealer.
13
Shutdown
2.1.2
Safety instructions – Assembly
If you do not intend to use the Casalift for an extended period, prepare it in the
following way:
Risk of trapping! Risk of crushing!
Remove dirt and dust.
There is an increased risk of trapping and crushing at all joints of the Casalift.
Full charge the battery.
While assembling the product, pay particular attention to these danger
points.
Care during the product service life:
Charge the battery once a month (see page 28).
Casalift 140E / 180E – 15.10.2010
Casalift 140E / 180E – 15.10.2010
Casalift 180 - SEPTEMBER 2017
Drive Devilbiss
Remedy
Unlock the EMERGENCY STOP
button by turning it to the right.
Plug the connectors into the con-
trol box properly.
Charge battery.
Replace the battery.
Replace lifting motor.
Replace control module.
Replace operating controls.
If you can clearly hear a click in
the control module when press-
ing and releasing the button, a
defect in the operating control
can be excluded.
Possible cause of fault: control
module defective!
Replace lifting motor.
Replace spreading motor
Replace control module.
GB – 35