Maintenance And Inspection; Inspection Intervals - DeVilbiss Healthcare drive Casalift 18001 Manuel D'utilisation

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Foreword
Disinfection
Introduction
Maintenance – Daily
1
9
Foreword
Disinfection
If disinfection is required during use, only suitable agents should be used.
1.1
Introduction
The Casalift 180 is a lightweight and versatile lift device for use in patient and
10
Reuse
geriatric care situations. Use of the lift is facilitated considerably by its spreading
The Casalift is intended for reuse. Before such reuse, it must be cleaned and dis-
travel leg design. After use, or for transport by car, it can be dismantled.
infected.
This manual was produced based upon the information available on the design
and operation of the Casalift which was available at time of printing. We reserve
11

Maintenance and inspection

the right to make changes due to technical advances.
Should you have any questions concerning the Casalift, please contact your spe-
Danger due to electrical and mechanical faults!
cialist Days Healthcare dealer (see dealer's stamp on the back of the envelope).
Immediately discontinue use of the Casalift if it is faulty, and secure it to pre-
vent unauthorised use.
1.2
Intended use
Report any defects which are discovered immediately to the appropriate par-
The Casalift 180 should be use exclusively for the purpose of lifting and moving
ties!
persons in need of care. It is only designed to move patients over short
distances.
11.1
Maintenance – Daily
Improper use will lead to the exclusion of the manufacturer's liability.
The following routine inspections
The Casalift may only be used by persons who have received instruction from the
should be performed by nursing staff
owner, and under consideration of the condition of the patient being transport-
or those persons responsible before
ed. In borderline cases, or if in doubt, consult the attending doctor.
every use:
Visual inspection of all compo-
Restriction on use!
nents of the Casalift for damage.
Patients who are heavier than 180 kg must not be lifted with the Casalift.
Check function of the parking
brakes.
1.3
Information on the operating manual
Check function of the lifting mo-
tor:
1.3.1
Symbols used in the operating manual
Lift (A)
The following symbols are used in the operating manual to identify particular
Lower (B)
hazards in using the product or information to simplify usage.
11.2
Maintenance – Annually
Caution!
This symbol identifies safety instructions which highlight risks in using the prod-
11.2.1 Requirements of inspection personnel
uct.
All inspection work described in the following may only be carried out by expe-
rienced personnel. The evaluation of the inspection list and servicing of the
Caution!
Casalift may only be performed by experienced personnel in possession of the
This symbol identifies safety instructions which highlight risks of crushing and
appropriate training, knowledge and experience.
trapping when using the product.
GB – 20
GB – 32
34
A
B
C
D
Check function of the spreading
motor:
Spread (C)
Reset (D)
Casalift 140E / 180E – 15.10.2010
Casalift 140E / 180E – 15.10.2010
Casalift 180 - SEPTEMBER 2017
Maintenance and inspection
Safety
Maintenance – Annually
Safety instructions

11.2.2 Inspection intervals

Inspection intervals depend upon the type of use the Casalift is subjected to.
Note
Under normal environmental and operating conditions, Days Healthcare recom-
Under this symbol, you can find helpful information on using the product.
mends inspection intervals of 1 year for inspecting the mechanical and electrical
components, and for the metrological control according to BGV A3.
1.3.2
Information on the operating manual
When the lift comes to be used again, the inspections should take place be-
Carefully read through the operating manual before assembling the Casalift and
fore each reuse (after assembly), and then annually.
bring it into service!
The inspection intervals should be shortened if the Casalift is subjected to an
Make sure that:
increased load (e.g. in nursing homes).
all persons who are to use the Casalift have read the operating manual.
In case of any doubt, please contact your specialist Days Healthcare dealer
(see dealer's stamp on the back of the envelope).
all persons charged with setting up or maintaining the Casalift always have
access to the operating manual.
11.2.3 Inspection report (template)
Damage caused through a failure to follow the information in the operating
manual is excluded from any and all warranty claims.
Description
The illustrations which appear in this operating manual are only examples of
Component
Inspection type
how to use Days Healthcare patient lifts, and may differ from your actual model.
Operating controls
Visual inspection of cable and plug
2
Safety
Electrical function
Lifting motor
Visual inspection of cable and plug
2.1
Safety instructions
Electrical function
Electrical emergency lowering function
2.1.1
Safety instructions – General
Mechanical emergency lowering function
Charging cable
Visual inspection for damage
Risk of accident!
Mechanical spread-
Joints free of play
Only use the Casalift for its intended use. The Casalift is not a piece of gym-
ing mechanism
Operates smoothly
nastics or playground equipment.
Electrical spreading
Joints free of play
Only use the Casalift if it is in a technically sound condition.
mechanism
Operates smoothly
Only use slings, sling seats or reclined slings which are in perfect condition. Replace
Control module
EMERGENCY STOP function
sling systems which are damaged or heavily worn immediately!
Housing undamaged
All faults which have a bearing on the function and safety of the Casalift
must be rectified without delay by a specialist dealer.
Proper fastening to the support
Battery charger
Measure voltage
2.1.2
Safety instructions – Assembly
Replace battery every 4 years
Support
Visual inspection of welds
Risk of trapping! Risk of crushing!
No damage
There is an increased risk of trapping and crushing at all joints of the Casalift.
Securely fitted into the chassis
While assembling the product, pay particular attention to these danger
Boom
Visual inspection of welds
points.
No damage
Joints free of play
Casalift 140E / 180E – 15.10.2010
Casalift 140E / 180E – 15.10.2010
Casalift 180 - SEPTEMBER 2017
Drive Devilbiss recom-
Drive Devilbiss dealer
Evaluation
OK
Defective
GB – 33
GB – 21
35

Publicité

Table des Matières
loading

Table des Matières