9 Error messages
If an error is detected, a corresponding error message appears that must be acknowledged
by pressing any key.
Possible error messages:
A data error was detected in the saved settings. This could be due to strong
E.10
electromagnetic interference or a component fault.
Press
E
Action:
restored for all adjustable parameters and all minimum and maximum values
will be deleted. Enter the data again from the beginning.
An error was detected in the saved calibration data. This could be due to
E.12
strong electromagnetic interference or a component fault.
Disconnect then reconnect the supply voltage to the instrument. If the error
Action:
persists, the instrument must be returned to the factory for recalibration or
repair.
A communication error was detected within the instrument. This could be due
E.21
to strong electromagnetic interference or a component fault.
Press
E
Action:
to the instrument. If the error still persists, please contact our service
department.
Status: 04.04.2018
and confirm "RES" by pressing "Yes". The factory settings will be
. If the error persists, disconnect then reconnect the supply voltage
HYDAC ELECTRONIC GMBH
15
E
Part No.: 669758