What if....?
Symbol
Causes
Electronic defect in the handpiece
or battery
Apparatus is overheated or undercooled
Electronic defect in the battery
Battery empty
Charging contacts contaminated
No display
during
charging
Battery in safety mode
LED of the
Power pack not connected or defective
charging
base is not
illuminated
14
Error Rectification
Remove and reinsert the battery. If the error remains,
replace the battery with the power pack (Click & Cure).
If the error remains, please contact your dealer or your
local Service Center.
Allow the apparatus to cool down (or assume room
temperature if it is undercooled) and try again after a
certain time. If the error remains, please contact your
dealer or your local Service Center.
Remove and reinsert the battery. If the error remains,
place the apparatus into the charging base.
If the error remains nonetheless, replace the battery
with the power pack (Click & Cure).
Please contact your dealer or your local Service Center.
Place the handpiece into the charging base. The corre-
sponding message ("Charging") will be displayed. If
the battery is not charged, the contacts have to be
cleaned. If the error remains, please contact your deal-
er or your local Service Center.
Check if the handpiece has been correctly placed into
the charging base. Clean battery contacts.
If cleaning the battery contacts does not improve the
results, remove the battery from the apparatus. Charge
the battery separately from the handpiece in the
charging base.
If the error remains, please contact your dealer or your
local Service Center.
Check if the power pack is correctly positioned in the
charging base or if the power pack is connected to the
power supply by means of the power cord (display on
the power pack lights up in green if it works correctly).
If the error remains, please contact your dealer or your
local Service Center.