What If; Error Rectification - Ivoclar Vivadent Bluephase Style M8 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 26

6. What if ...?

Indicator
Causes
The device is overheated.
all LEDs
orange
Electronic component of the handpiece
all LEDs
red
defective
– Power pack not connected or defective
The charging
– Battery fully charged
base is not illu-
minated during
charging
Battery empty
No indicator
activity of the
device with the
battery in place
Battery contacts contaminated
12

Error rectification

Allow the device to cool down and try again after
a certain time. If the error reappears, please contact
your dealer or your local Service Center.
Remove and reinsert the battery. If the error reap-
pears, please contact your dealer or your local Service
Center.
Check if the power pack is correctly positioned in
the charging base or if the power pack is connected
to the power supply by means of the power cord
(display on the power pack lights up in green if it
works correctly).
Place the device in the charging base and charge
for at least 2 hours.
Remove battery and clean the battery contacts.

Publicité

Table des Matières
loading

Table des Matières