Technical Characteristics - Gima SUPER TOBI Mode D'emploi

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

TECHNICAL CHARACTERISTICS

TYPOLOGY (MDD 93/42/EEC)
MODEL
UNI EN ISO 10079-1
POWER FEEDING
POWER CONSUMPTION
FUSE
MAXIMUM SUCTION PRESSURE (without jar)
MINIMUM SUCTION PRESSURE (without jar)
MAXIMUM SUCTION FLOW (without jar)
WEIGHT
SIZE
DUTY CYCLE
(to 35°C and 110% operating voltage)
WORKING CONDITION
CONSERVATION CONDITION AND TRASPORT
~
Hz
I
O
Dispositivo Medico Classe IIa
SUPER TOBI
HIGH VACUUM / HIGH FLOW
230V~ / 50Hz
107 VA
F 1 x 1.6A 250V
-80kPa (-0.80 Bar)
Less – 40kPa (-0.40 bar)
40 l/min
3.6Kg
370 x 220 x 210mm
120 min ON / 60 min OFF
Room temperature:
Room humidity percentage:
Altitude:
Room temperature:
Room humidity percentage:
SYMBOLS
Class II isolation equipment
CE marking in conformity with EC directive 93/42/EEC
Manufacturer by CA.MI. di Attolini Mario & C. s.n.c.
Via Ugo La Malfa 31 – 43010 Pilastro (PR) Italy
Warning, consult the instruction manual
To Preserve in place coolness and dry land
Conservation temperature: -40 ÷ 70°C
Type B equipment
Fuse
Alternate Current
Mains Frequency
ON
OFF
9
110V~ / 60Hz
(no CE 0123)
165 VA
F 1 x 4A 250V
10 ÷ 40°C
20 ÷ 85% RH
0 ÷ 2000m s.l.m.
-40÷ 70°C
10 ÷ 95% RH

Publicité

Table des Matières
loading

Table des Matières