2
93 °C
For cleaning of the tubing see chapter 5.7.
Clean and disinfect the remaining parts in a
mechanical washer-disinfector. Use
appropriate inserts for the disinfection of small
parts. Use a common cleaning detergent
for washer-disinfectors (NTA-, phosphate-,
chlorine- and tenside-free, alkaline with pH
level 9,5–12). Disinfect for five minutes at a
temperature of 93 °C. Select an appropriate
program, according to the above instructions,
that includes drying*.
5.9 Reusable pump set – Sterilisation
If national, regional or in-house regulations
require sterilisation, do as follows:
I Check that the parts of the pump set are
clean, dry and intact.
I Wrap all parts, including the membrane and
the membrane cap, in appropriate sterile
packaging, such as a paper film.
3
Storage
Put the clean parts in a clean storage
bag or a clean environment. It is
important that all residual moisture
dries. Do not store parts in an airtight
container/bag.
*If the appropriate program does not include
drying, dry the clean parts with a clean cloth
or leave to dry on a clean cloth.
I Sterilise the product in the autoclave at
134 °C for 3 minutes. To avoid damage,
do not stack or apply external pressure.
I Leave the product parts in the sterile
packaging and store them in a clean, dry
and dust-free environment. It is important
that all residual moisture dries. Do not store
in an airtight container/bag.
25