Removing the power plug adapter
Press on the ribbed button from
t
above while sliding the power plug
adapter backward.
Push the power plug adapter out of
t
the AC adapter and remove it.
Connecting the AC Adapter
Check the voltage rating on the AC adapter's type plate. Make
t
sure that the voltage rating printed on this unit matches the
local supply voltage at the place of installation.
If the stated supply voltage does not comply with the local sup-
t
ply voltage or there is no suitable AC adapter available: Do not
use the AC adapter. Contact Sartorius Service.
Only use original Sartorius AC adapters.
t
Connect the angle plug (2) to the
t
1
device's power supply socket (operat-
ing voltage connection).
Connect the power cable (1) to the
t
wall outlet (supply voltage) at the
installation location.
2
Warm-up Time
0
y To ensure accurate results are delivered,
the balance must warm up for at least
30 minutes after initial connection to the
power supply. Only after this time will the
device have reached the required operating
temperature.
30
When a verified balance of accuracy class k for use in
legal metrology is connected to the mains power, it must
warm up for at least one hour before operation.
Leveling
h
Leveling the balance compensates for slant or unevenness
at the place of installation. The balance must be perfectly
horizontal to ensure consistent, reproducible weighing
results.
Note: The balance needs to be re-leveled and adjusted
each time its setup location is changed.
6
Installation and Operating Instructions Secura
Models Secura
When "Level" appears on the display, you must level the balance.
Models Quintix
Instructions for Use in Legal Metrology
Additional Display Elements:
– Metrology line with additional information about minimum load
(Min) and verification scale interval (e)
Interface Specification (Data Output)
Settings „PC – SBI" and „USB RS232 SBI" for the identification of
digits not verified for use in legal metrology: In the setting „SBI"
there is no automatic identification of display digits not verified for
use in legal metrology. Corresponding measures or settings must be
carried out on the peripheral device.
Below-balance Weighing
A port for a below-balance weighing hook is located on the bottom
of the balance.
h
®
®
®
, Quintix
, Practum
®
:
Make sure that the weighing pan is
t
unloaded.
Touch "LEVEL!"
t
Follow the instructions on the
t
interactive display.
®
®
and Practum
:
Turn the two feet as shown in the
t
illustration until the air bubble is
centered within the circle of the level
indicator.
– Air bubble at 12 o'clock:
turn the two feet clockwise.
– Air bubble at 3 o'clock: turn the left
foot clockwise and the right foot
counterclockwise.
– Air bubble at 6 o'clock:
turn the two feet counterclockwise.
– Air bubble at 9 o'clock: turn the left
foot counterclockwise and the right
foot clockwise.
Warning symbol to indicate calculated values, negative
gross values or values that are smaller than the minimum
sample quantity (SQmin).
Setup Menu „Settings": On balances verified for use in
legal metrology, not all functions and settings are available.
Install a draft protection shield.