Operation

7.3.22 "LANGUAG." Menu

Parameter
Setting values
Language
ENGLISH,
DEUTSCH,
FRANC .
ITAL ., ESPANOL,
PORTUG.,
РYССKИИ,
POLSKI
* Factory setting
8
Operation
8.1

Switching the Device On and Off

Requirements
The device is connected to the power supply .
Procedure
CAUTION Pointed or sharp-edged objects may damage
t
the operating display!
Only touch the operating display with your fingertips .
t
To switch the device on: Press the [On/Off] button .
t
To switch the device off: Press and hold the [On/Off]
t
button .
8.2

Waiting for the Warm-up Time

In order to provide accurate weighing results, the device
must have reached the necessary operating temperature .
The operating temperature is reached after the warm-up time .
The warm-up time starts when the device is switched on .
Procedure
Switch on the device .
t
Wait until the operating temperature is reached (warm-up
t
time see Chapter "15 .7 Warm-up Time", page 37) .
If a weighing process is carried out on a conformity-
assessed device during the warm-up time:
The weight value is marked as invalid .
8.3

Leveling the Device with a Level

Unevenness at the installation site of the device may result in
incorrect weighing results . Leveling balances out unevenness at
the installation site .
28
Entris
®
BCE Operating Instructions
Explanation
Defines the menu language .
Procedure
In order to level the device: Bring the air bubble into the
t
middle of the circular marking . To do so, turn the leveling
feet to the left or right .
8.4

Overview of Calibration and Adjustment

During calibration, a calibration weight is used to determine
the deviation of the displayed value from the actual value .
The subsequent adjustment eliminates this deviation .
We recommend regular calibration and adjustment:
− Daily, every time the device is switched on
− After every leveling
− After changing the ambient conditions (temperature,
humidity, or air pressure)
− After setting the device up at a new installation site
8.5
Calibrating and Adjusting Device with
isoCAL Function (Only Model i-1x)
Triggers for the automatic start of the isoCAL function are:
− The ambient temperature has changed since the last
adjustment process .
− The interval time was exceeded (interval time, see Chapter
"15 .7 Warm-up Time", page 37) .
If this relates to a conformity-assessed model:
The device has been disconnected from the power
supply since the last adjustment .
Requirements
− The automatic or manual start of the isoCal function is set
in the menu (see Chapter "7 .2 .1 Switching the isoCAL
Function On or Off (Only Model i-1x)", page 18) .
− The device is not located in the menu settings .
− The load on the weighing pan remains unchanged for
2 minutes .
− The load on the scales amounts to no more than 2% of the
maximum load .
− The device does not register an input for 2 minutes .
Procedure
If the automatic start of the isoCAL function is triggered:
t
The [isoCAL] button flashes in the operating display .
y
The operating display shows "CAL ." for 19 seconds .
y
If no load change or no operation takes place on the
y
device before the expiration of the time display:
The isoCAL function starts .
If the manual start of the isoCAL function is triggered:
t
The [isoCAL] button flashes in the operating display .
y
Press the [isoCAL] button .
t
The isoCAL function starts .