7. Assembling your breast pump
Only use clean and dry parts for assembly. Refer to chapter 6 for cleaning.
7.1 Selecting the correct breast shield size
Make sure to use the correct breast shield size. It is an essential component of effective
16
pumping, helping to optimize milk flow.
Refer to figure ¢6 on the foldout pages.
1. Select the breast shield size by measuring the diameter of your nipple ( ¢6-01).
2. Select the breast shield size that is slightly larger than your nipple ( ¢6-02).
21mm (without the O-ring) and 24mm (already applied with the O-ring) size breast shields
are included in the pack. 27mm size breast shield is available for purchase.
See chapter 13 for ordering information.
For more information on breast shield sizing and other available sizes, visit www.medela.com
or see chapter 13 for ordering information.
7.2 Assembling the collection cups
Only use clean and dry parts for assembly. Refer to chapter 6 for cleaning.
Parts needed:
• Tubing
• Breast shield with O-ring
• Motor unit
CAUTION
To prevent contamination of your milk:
• Wash hands thoroughly with soap and water before touching the motor unit, parts of the
collection cups and breasts.
• Dry your hands with a fresh towel or a single-use paper towel.
• Avoid touching the inside of the collection cups components.
To make sure that your breast pump works properly and safely:
• Check collection cups components for wear or damage before use. Replace if necessary.
• Always inspect all parts prior to use for cleanliness. If dirty, clean according to chapter 6.
• Only use genuine Medela parts. See chapter 13 for details.
NOTICE
To prevent damage to the breast pump all components must be completely dry
before use.
Refer to figure ¢7 on the foldout pages.
1. ¢7-01 Wash hands thoroughly with soap and water before touching the motor unit,
parts of the collection cups and breasts. Dry your hands with a clean towel.
2. ¢7-02 Confirm that the O-ring is applied to the breast shield.
3. ¢7-03 First assemble the valve of the yellow membrane on the front of the breast shield.
4. ¢7-04 Align the disc of the yellow membrane on the designated circular slot using the
alignment tab at the top to ensure that it is at the right place. Press around the edges to
make sure it is fully assembled.
15400140_A_0523-Freestyle-Hands-free-MD-IFU-EN-FR-ES_052323.indb 16
15400140_A_0523-Freestyle-Hands-free-MD-IFU-EN-FR-ES_052323.indb 16
• Yellow membrane
• Outer shell