USER CALIBRATION
Use this calibration method to calibrate the balance using
a defined mass.
1.
With no load on the scale, press USER CALIBRATION
from the calibration sub-menu. The display shows
the current full load (span) calibration point and
calibration unit of measure.
2. To change the value to a user-defined weight,
touch ENTER VALUE. Enter the weight via the pop-up
screen. The balance then prompts for the user-
defined weight.
3. Place the specified calibration weight on the
balance and the calibration process completes.
The display shows status, then returns to the current
application.
CALIBRATION TEST
Use this calibration method to compare a known
calibration weight against the stored span calibration
data.
1.
With no load on the balance, press Cal Test to
begin the process.
2. The balance captures the zero point, then prompts
for the span value.
3. The display shows status, followed by the difference
between the current calibration weight and the
stored calibration data.
4. After three seconds, the display returns to the current
application.
TURNING AN APPLICATION ON/OFF
1.
In the balance Setup Menu, enter the Application
Modes sub-menu to activate the desired balance
applications. Only one can be running at a time.
2. Touch the line of the desired application.
3. In the item option screen, touch ENABLED to toggle
it on or off.
PAGE 6 OF 106
OPERATION CONTINUED
4. Once an application is enabled (turned on), it may
5. The current menu item status is shown:
APPLICATIONS
The balance can be configured to operate in various
application modes. Touch APPLICATIONS to choose – or
the top button in the application area.
be chosen from the Applications shortcut (top-left of
all home screens).
OFF = disabled
ON = enabled
NOTE: Before using any application, ensure the
balance has been leveled and calibrated. The
balance is shipped with all units enabled.