DENSITY DETERMINATION
The balance can be used to determine an object's density.
Four types of density determination can be made:
• Solids more dense than water
• Solids less dense than water
• Liquid density
• Porous material (impregnated with oil)
A density determination kit is designed to be used
with the balance. Illustrations in this procedure refer
to the density kit. However, you may use whatever
lab apparatus that suits the requirements for density
measurements. The balance software includes a built-in
reference density table for water at temperatures
between 50 ° F and 86 ° F. Review this entire section
before attempting density measurements.
When making density measurements the material
should weigh at least 10 mg on an analytical balance
and 100 mg on a precision balance.
Density can be determined manually (with a key
press after each step) or automatically (weights are
automatically recorded). Before making density
measurements, establish the Application Settings.
MEASURING THE DENSITY OF A SOLID USING
WATER (DEFAULT)
1.
Press the SETUP icon to open the density
determination settings screen.
2. Confirm the following setups are selected:
• Density type: Solid
• Liquid type: Water
• Porous material: Off
3. Press DONE to return to the density determination
home screen.
PAGE 22 OF 106
OPERATION CONTINUED
EDIT SETTINGS SCREEN
• Settings available: Water temperature
• Functions available: Return to application
1. To adjust the water temperature value, touch the
2. A numeric input window appears. Enter the actual
3. The display returns to the previous screen with the
4. To begin the density determination process, press
DENSITY DETERMINATION STEPS
1.
2. Weigh the sample submerged in the liquid. Follow
WATER TEMP button. The balance calculates water
density based on the water temperature value
entered (internal lookup table). Measure the actual
water temperature using a precision thermometer.
water temperature, then press DONE.
new value highlighted. To return to the density
determination home screen, touch RETURN TO
APPLICATION.
BEGIN DENSITY DETERMINATION.
Weigh the sample in air. Press BEGIN DENSITY
CALCULATION. Follow the screen instructions, then
press ACCEPT to store the dry sample weight ("in air").
the screen instructions, then press ACCEPT to store
the sample weight (submerged in liquid).