6. To adjust the water temperature, touch the
WATER TEMP button.
7. The balance calculates water density based on
the water temperature value entered. Measure
the actual water temperature using a precision
thermometer.
8. A numeric input window is displayed.
9. Key in the desired temperature, then press √.
10. The display returns to the previous screen with the
new value highlighted.
11. Continue to enter the required information following
the above procedure.
12. When finished, press RETURN TO APPLICATION. The
application home screen appears, and you can
begin the pipette adjustment process.
BEGIN PIPETTE ADJUSTMENT PROCESS
1. Touch BEGIN PIPETTE ADJUSTMENT.
2. Follow the screen instructions, to place container on
the pan. Press TARE.
PAGE 28 OF 106
OPERATION CONTINUED
DISPENSE THE FIRST SAMPLE
1.
2. Continue to follow the screen instructions by
3. Once all the sample weights have been dispensed,
4. Press GRAPH to see results plotted. Press RESULT to
5. To view either the results or graph screens again
Follow the screen instructions, then press ACCEPT to
store the first sample weight (liquid).
dispensing samples and pressing ACCEPT to
store each sample weight (liquid).
the results screen is automatically displayed.
NOTE: The default number of samples is 10 but
can be changed in Setup from 2 to 30 samples.
return to the result screen. Press CLOSE to return to
the pipette adjustment home screen.
from the home screen, press the VIEW RESULT button.
To start a new pipette adjustment process, press
BEGIN PIPETTE ADJUSTMENT.