11. To clear all items, press CLEAR ALL. A confirmation
is required. When finished, press RETURN TO
APPLICATION.
12. Once setup is complete and the application screen
is displayed, follow the screen instructions.
13. Add the first item then press ACCEPT to store the
value in memory. The weight of the item is recorded,
then the screen prompts for the next component.
Unit Cost, Total Weight and Total Cost are also
displayed.
14. Continue to follow the screen instructions to add the
items to the pan and touch the ACCEPT button until
each item has been weighed.
15. Press the VIEW DATA button at any time to show the
current results: Item, Item Cost and Item Weight.
16. After the last item has been weighted, press
VIEW DATA to see the results.
NOTE: Cancel the current data set by pressing
CLEAR DATA.
INGREDIENT COSTING RESULTS
1.
Press PRINT to send the data to the enabled COM
port.
2. Press CLOSE to close the results window and return
to the Application home screen.
PAGE 27 OF 106
OPERATION CONTINUED
PIPETTE ADJUSTMENT
Pipette adjustment checks the accuracy and precision
values of pipettes by weight analysis. The balance has a
built-in density reference table for water at temperatures
between 50 ° F and 86 ° F. If other liquids are used for
pipette calibration, enter the liquid's density at current
room temperature in g/cm
made within the balance, the atmospheric pressure is
also required.
Pipette adjustment can be accomplished manually (with
a key press after each step) or automatically (weights are
automatically recorded when stable). Calculations made
within the balance provide results of inaccuracy and
imprecision. Up to 30 samples can be used.
PIPETTE ADJUSTMENT – MANUAL (DEFAULT)
1. Confirm the default values displayed are correct:
• Nominal, Inaccuracy, Imprecision, Density,
2. To edit the default values, touch the EDIT SETTINGS
• Settings available: NOMINAL CAPACITY, WATER
• Functions available: RETURN TO APPLICATION
3. To adjust the pipette nominal capacity value, touch
4. Enter the nominal capacity in the keyboard screen
5. The display returns to the previous screen with the
Atmospheric Pressure, Water Temperature, etc.
button.
TEMP, INACCURACY, ATMOSPHERIC PRESS,
IMPRECISION, PIPETTE NAME, PIPETTE NUMBER
the NOMINAL button.
that appears, and press √.
new value highlighted.
NOTE: Capacity can be measured in mL or µL,
depending on the application setup.
. Since all calculations are
3
0924 IH-4737