Alarm message linked to the FlowSense® sensor
During ventilation, if the FlowSense® sensor is defective or not properly connected or if a foreign object on the
connector interferes with the signal transmission, an alarm screen is displayed:
You must reconnect a functional FlowSense® sensor as described above to clear the "FlowSense failure" alarm
Note
The "FlowSense failure" alarm screen can only be displayed from the main screen
EOlife® device failure
If there is an EOlife® device failure, the red LED in the top left corner of the screen comes on This red LED tells you
when there is a device failure whether it is on or off
Warning
Do not use the EOlife® device if the red LED stays on
See "X What to do in the event of an incident"
d) Standby mode
If EOlife® remains in standby mode for more than 1 hour, the device will turn itself off completely
2. EOlife® device set-up screen
a) Set-up screen display
1.
Start up EOlife® by pressing the physical ON/OFF button
2. Press normally the physical ON/OFF button to show the summary screen
FlowSense failure
Alarm screen
"FlowSense failure"
Summary screen
41
EN