A.7
Installation of video camera
with Lipstick interface
Installation of video camera head
Affix the adapter for Lipstick to C-mount. Only tighten the 3
set screws sufficiently enough to enable the camera adapter
still to be turned.
Image position on the monitor
Turn the camera head until the image on the monitor is correctly
margin-adjusted.
Image focus on the monitor
Use the focus test rod. Focus a slit of 8 mm height and 4 mm
width onto the focus test rod. Shift the illumination to 45° from
the microscope.
Set a 16x magnification. Adjust the video camera depth. The
area, in which the structure of the focus rod surface is seen
sharply in focus, should lie in the centre of the light slit (107),
whereas the structure to the left and right is shown out of
focus (108).
The 3 set screws are still slightly loose and the camera height
can be adjusted vertically by turning the camera adapter - but
not the camera which can be held with one hand so that it
cannot turn. When the image on the video monitor screen
appears sharply in focus, the lock nut of the camera can be
firmly tightened.
The position of the image on the monitor as to margin correctness
should be checked again and correctly adjusted.The camera
adapter can then be secured to the microscope housing by
tightening the 3 set screws.
CAUTION
Only switch on the camera control unit when all
components have been correctly installed.
Observe the instruction manual for the video
camera when making these connections.
A.8
Mounting of headrest and power supply
The corresponding Haag-Streit operating instructions should
be observed!
107
108
1500.7200631.04040
Adapter für Lipstick auf C-mount
Adaptateur pour mini caméra sur monture C
Adapter for Lipstick on C-mount
45°
© Haag-Streit AG, CH-3098 Koeniz, Switzerland
35