Safety
Ambient conditions
• Transportation Temperature
Air pressure
Relative humidity 10%
• Storage
Temperature
Air pressure
Relative humidity 10%
• Working
Temperature
Air pressure
Relative humidity 30%
Installation
• To avoid condensation, allow the instrument to adjust to room
temperature for several hours before unpacking.
• Always mount the slit lamp and the headrest on an electric
insulated and fire resistant table top.
• The rail covers (30) prevent from a tip over of the slit lamp.
• Check: do the connection pieces of the accessories fit (like
screwed connections and bayonet fitting)?
Operation and surrounding
• Only qualified and trained personnel should operate the equip-
ment, the training is at the owner's responsibility.
• Use only Haag-Streit accessories.
• Shut down after every use. In case the dust cover is used:
risk of overheating.
• Never use the ocular to look at the sun.
• Never operate the instrument in the same room with combu-
stible gases, volatile solvents (alcohol, benzol) or flammable
anesthetic agents.
• Avoid humidity.
Changing the light bulbs
• Switch off the main switch - disconnect the mains connector.
• Allow the bulb to cool down.
• Do not open the slit lamp for any other reason than to change
the light bulb.
Electrical
• Only a hospital grade 3-conductor electrical power supply
cable must be used. For USA and Canada: Detachable Power
Supply Cord Set, UL Listed, type SJE, SJT or SJO, 3-conduc-
tor, not smaller than 18 AWG.
• Plug, cable and ground lead connection of the socket have
to be in perfect condition.
Cleaning
• Clean the slit lamp housing only with a slightly water dam-
pened cloth.
• No liquids or corrosive agents.
• Clean only the exterior surfaces of the optical parts.
Light toxicity
• Long and intensive illumination could damage the retina. The
brightness of illumination should be kept at a minimum, to view
the structure of the target clearly. To avoid potential eye dama-
ge, the examination should not last longer than necessary.
-40 °C
to +70 °C
500 hPa to 1060 hPa
to 95%
-10 °C
to +55 °C
700 hPa to 1060 hPa
to 95%
+10 °C to +35 °C
800 hPa to 1060 hPa
to 75%
30
Instrumententisch HSM-901 Typ M
Unbedingt diesen Sicherheitshinweis beachten
Table d' instruments HSM-901 type M
Faire attention aux indications de sécurité suivantes
Instrument Table HSM-901 type M
Observe the following safety precautions
1500.7200631.04040
Höhenverstellung nur mit aufgesetztem Gerät betätigen
Do not use height adjustment without instrument in place
Ne pas utiliser le réglage en hauteur quand l'instrument n'est pas en place
Patents
US Pat. 3.685.779
Haag-Streit BERN
and other countries
SWISS MADE
© Haag-Streit AG, CH-3098 Koeniz, Switzerland
11