The Feed Reduction Feature
26
Unitom-50
Instruction Manual
On Unitom-50, the desired Feed Speed is pre-set and kept
constant throughout the cutting process.
If the pre-set Feed Speed is too high in relation to the hardness of
the material or size of the sample, Unitom-50 will automatically
reduce the Feed Speed (default setting).
The display shows the following text:
Unitom-50 continues to cut at the reduced Feed Speed and
reduces the Feed Speed repeatedly, if necessary.
Next time the same type of sample has to be cut, the reduced
value for Feed Speed can be used.
After the Feed Reduction Feature has been in operation, the
display shows the following, next time START
If the same type of sample has to be cut, press F1.
If Feed Reduction is not desired, feed reduction is set to Off. This
can be used to force a dressing of the cut-off wheel. However, this
might result in the motor stopping due to overload.
s
is pressed: