Instruction Manual
The cooling pump is delivered wrapped in corrugated cardboard
and placed inside the cutting chamber. The pump is connected as
follows:
Remove the cooling pump from its wrapping.
Mount the inlet hose on the pump using a hose clamp.
Mount the other end of the inlet hose on the back of Unitom-50
using a hose clamp.
Lead the pump through the opening in the back of the table.
Mount the drain elbow pipe on the water outlet tube in the
hose box at the bottom of Unitom-50.
Mount the outlet hose with the hose clamp on the drain elbow
pipe. Adjust the length of the hose, so as to direct the water
into the recirculation tank.
Place the tank on the trolley.
Place a disposable plastic insert (EXOSP, ordered separately)
in the tank and fold it over the edge.
Fill the tank with 63 l of water and 2 l Struers Additive for
cooling fluid. The water level should be 8-10 cm below the
upper edge of the tank.
A too high level of coolant in the tank might damage the pump. To avoid
this: place the disposable insert so that the pre-punched hole is in front of
the overflow aperture in the tank.
Always maintain the correct concentration of Struers Additive in the cooling
water (percentage stated on the container of the Additive). Remember to
add Struers Additive for cooling fluid each time you refill with water.
IMPORTANT