mainTenance, TrOuBleshOOTing

10.1.

maintenance work

The flow transmitter / pulse divider Type 8022 is maintenance-free when operated according to these operating
instructions.
10.2.

error messages

Error messages are only displayed at the Read level. They are shown alternately (flashing) with the process value.
error
cause
ERR1
Value cannot be displayed
(e.g. value too high).
ERR2
Input frequency of sensor higher than 600 Hz.
ERR3
When operating as a flow transmitter:
Calculated output current not within range of
4-20 mA or K factor = 0.
When operating as a pulse divider:
Calculated output frequency higher than 600 Hz
or K factor = 0.
ERR4
Limit values of the flow rate range associated to
the 4-20 mA current output not correct (Low >
High).
Table 2:
Error messages
10.3.

Default values

Upon delivery, the following default values are saved:
parameter
Unit
K factor (of fitting used)
K factor (pulse dividing)
Output signal (OUT)
Lower flow rate limit
Upper flow rate limit
Table 3:
Default values
24
English
Type 8022
Maintenance,Troubleshooting
troubleshooting
• Change the fow rate unit
(see chapter 9.4.1. Unit – Setting the unit for
the flow rate).
• Use a suitable sensor.
• Correctly set the flow rate range associated to
the 4-20 mA current output.
• Use a different sensor and/or correctly set the
K factor of the fitting used.
• Change the division ratio
(setting of pulse dividing K factor).
• Use a different sensor and/or correctly set the
K factor of the fitting used.
Correctly set the values.
Value
Liters/second [l/s]
1 pulse/liter [imp/l]
1 pulse/liter [imp/l]
4-20 mA
0 liters/second [l/s]
250 liters/second [l/s]