Connection A - High Vacuum Flange
For ISO-K flange connections, fix the two flanges with the clamps as shown in
the figure.
Figure 51
WARNING!
It is mandatory to connect the pump in such a way to withstand the torque
specified in the "IN-STRUCTION FOR USE" section. Specifically pay attention
to clamp design, material of clamps and bolts and bolt fixing torque
For ConFlat flange connections we recommend using Agilent hardware.
To facilitate assembly and dismantling, apply Fel-pro C-100 high temperature
lubricant to the screw threads protruding from the flange and between the
nuts and flange.
Note that the connections can be made only with the bolt in the lower side.
Attach the units and tighten each one in turn. Re-peat the sequential tightening
until the flange faces meet.
TV 1001 Navigator User Manual / 87-900-945-01(L)
Technical Information
TV 1001 Controller Installation
17