Signal Description
Start/Stop: input signal to start or stop the pump. With the supplied cover
connector the START/STOP (+) signal is connected to the +24 Vdc pin and the
START/STOP (-) signal to the GROUND pin: in this condition the pump
automatically starts as soon as the controller recognises the input supply
("Plug & Pump").
Interlock: input signal to control the pump rotation. With the supplied cover
connector the interlock (+) signal is connected to the +24 Vdc pin and the
interlock (-) signal to the GROUND pin.
Speed setting: PWM input signal to set the pump speed. The PWM signal
characteristics must be the following:
Figure 14
The duty cycle percentage is referred to the low level portion of the PWM signal
NOTE
TV 1001 Navigator User Manual / 87-900-945-01(L)
frequency: 100 Hz +/-20 %
amplitude: 24 V max
duty cycle range: from 25 % to 75 %
corresponding to an output frequency from 272 Hz to 633 Hz (see the
following diagram):
Technical Information