Télécharger Imprimer la page

Agilent Technologies Navigator TV 1001 Notice De Mode D'emploi page 265

Publicité

J2 – Serial
Figure 22
This is a 9 pin D-type serial input/output connector to control via an RS 232 or
RS 485 connection the TV 1001.
Tab. 5
Pin N.
1
2
3
4
5
6
7
8
9
Note that the vent valve can also be controlled by means of the serial
connection.
A serial communication kit with a serial cable and the T-Plus software is
available (optional).
TV 1001 Navigator User Manual / 87-900-945-01(L)
Typical output connection
Signal name
SPARE
TX (RS232)
RX (RS232)
SPARE
GND
A + (RS485)
SPARE
B – (RS485)
RESERVED
Technical Information
Interconnections
265/310
17

Publicité

loading