Introduction
This manual is designed to assist with the operation and maintenance of the P100 Powered Support Surface. Carefully
read this manual thoroughly before using or beginning maintenance on the support surface. To ensure safe operation
of this equipment, it is recommended that methods and procedures are established for educating and training staff on
the safe operation of the support surface.
CONTRAINDICATIONS
Air support therapy is not recommended when spinal stability is a concern. This support surface is not intended to
support a patient in the prone position.
INTENDED USE OF PRODUCT
P100 is a constant low pressure powered support surface intended to provide pressure redistribution to aid in the
prevention and treatment of pressure ulcers. The system consists of a control unit combined with an alternating air
cell mattress. The air cells redistribute the patient's weight over the surface and aid in the reduction of tissue interface
pressure. It is recommended that the product be operated by personnel who are qualified to perform general nursing
procedures and have received adequate training in the prevention and treatment of pressure ulcers.
This support surface is intended to be used with human patients in a general hospital, nursing home or homecare
environment and for patients at risk of developing pressure ulcers, as well as those who require therapy for pre- existing
pressure ulcers. The safe working load for P100 is 135 kg /297 lbs; the patient must not exceed safe working load
specified by the support surface, frame, and accessories. Patients shall meet the minimum age requirement of 2 years
old.
P100 shall be used with a mattress cover at all times.
The support surface is not intended to be a sterile product nor is it intended to include a measuring function.
EXPECTED SERVICE LIFE
The products are intended to offer safe and reliable operation when in use or installed according to the instructions
provided by Stryker Medical. Stryker Medical recommends that the system be inspected and serviced by authorized
technicians if there are any signs of wear or concerns with device function and indication on products. Otherwise,
service and inspection of the devices generally should not be required. P100 has an expected service life of 2 years.
Return To Table of Contents
www.stryker.com
476004-5210 V1.5