(3)
Use control push buttons to select and change the required parameter.
Right button [ ]
Up Button [ ]
Left button [ ESC ]
Down button [ ]
(4)
Press [ENTER] to accept the parameter changes.
NOTE: Pressing [ESCAPE] returns unit to the upper menu.
(5)
Use control push buttons to:
Accept changes and leave menu
Do not accept changes and leave menu
Return to the System menu.
NOTE: Common relays are not configurable. Their condition is set to the following fixed settings:
Fixed Setting for Common Relays
•
Alarm 1 on any channel de-energizes common Alarm 1 relay on the Main board.
•
Alarm 2 on any channel de-energizes common Alarm 2 relay on the Main board.
•
Fault on any channel de-energizes Fault relay on the Main board.
•
System fault de-energizes Fault relay on the Main board.