Troubleshooting

Troubleshooting

Telemetry Help

The RC has an effective communication range of up to 91.4 cm (36 inches) with the Boston Scientific
rechargeable Stimulator, and a nominal effective communication range of 55.9 cm (22 inches) with the
Boston Scientific non-rechargeable Stimulator. Common sources of interference, such as televisions
and computer monitors, can affect wireless communication. You will continue to have access to the RC
menus even if the RC is outside its effective communication range. If you enter a command to change
your settings or programs while the remote control is outside of its effective communication range, you
will have several seconds to move the RC within range of the Stimulator to send the command. After
the command is received by the Stimulator, the remote control will make a double-beep then display the
requested change on the screen. If the RC is unable to communicate with the Stimulator, the Telemetry
Error message will appear:
To retry communication with the Stimulator – this resends the last command – select Retry
.
To cancel communication and return to the screen, press
.
The Telemetry Help tool is also available to assist you in improving communication between the RC and
?
the Stimulator. To use the Telemetry Help tool, select Telemetry Help
. The following screen will be
displayed:
Clinician Remote Control Directions for Use
90930936-14
23 of 125