electrician can approve the nursing bed for use after making the measurements and
performing maintenance.
Servicing must be confirmed in the service plan when at least the above-mentioned aspects have
been checked. Should your specialist dealer not perform the services you request, then contact the
manufacturer. We shall gladly refer you to authorized specialist dealers in your area.
The manufacturer is not liable for damage caused by the lack of or improper servicing.
Requirement:
1.) Visual check:
1.1) Power cable
EPR cable or comparable quality
1.2) Condition of power cable
In particular look for: Crushing, abrasion, cuts, visible
insulation of the inner conductor, visible metallic veins,
kinks, lumpiness, color changes of the outer sleeve,
brittle spots)
1.3) Status of internal wiring
In particular look for: Crushing, abrasion, cuts, visible
insulation of the inner conductor, visible metallic veins,
kinks, lumpiness, color changes of the outer sleeve,
brittle spots)
1.4) Sufficient ease of pulling and power cable kink
protection
1.5) Sufficient ease of pulling and kink protection for
internal cables
1.6) The power cable and other electric leads must be
wired securely so that shearing, squashing, and
other forms of mechanical stress are unlikely.
1.7) Moisture protection IP66
1.8) Inspecting device for suspended power cables
during transport.
1.9) Check the frame parts for plastic deformation and/or
wear and tear.
- bottom of bed
- bed lift
- backrest, seat area, foot rest section
- boom
- Fastening of boom
- Rolling castors
1.10) Housing
Visual inspection of all housings for damage. The
screws need to be tightly fastened. The seals must not
have any visual damage
Table (A) on Picture (B)
Remark:
n.a.
Yes
n.a.
Yes
n.a.
Yes
n.a.
Yes
n.a.
Yes
n.a.
Yes
n.a.
Yes
n.a.
Yes
n.a.
Yes
n.a.
Yes
n.a.
Yes