Install protective hood, (option P/N 591.00010.00)

This protective hood (25) protects the user from glass fragments, a possible shock wave
will escape via the top end. This hood guarantees adequate protection in both, the
operational mode and with the lift in upper position.
The protective hood (25) with its structural member is attached to the vertical stop (15)
with 2 screws. For opening or closing the protective hood (25) through 90 degrees, use
the handle provided. To displace the protective hood (25) loosen clamp screw (16); this
way, you can locate the hood over the center of the hot bath; if, for example (using bigger-
size evaporator flasks), the hot bath may be displaced to the right and the protective hood
(25) arranged in the right place. Secure with clamp screw (16).
fig. 13