Triggering
Command
Set commands
VAC=XXXX!
DP=XXX!
DT=XXX!
DV=XXX!
BATH=XXXX!
SPEED=XXX!
Actions
SVX!
VVX!
BVX!
MP!
MT!
SBX!
SSX!
LU!
LD!

32. Protective enclosure

The protective enclosure (1) is an extra
safety feature for the operator. Its large
pane of a special safety glass
guarantees a clear view of the rotating
flask. A handle (2) on the front is used
for closing and opening. In both
extreme positions, it is held by a
pneumatic strut.
Caution: Flask will not rotate unless
the hood is closed. Opening the
enclosure during operation will stop
rotation.
Safety Cover LED (6) illuminates.
After closing the enclosure, rotation is
restarted by the Start Stop key.
action
remarks
Set Vacuum
4-digit in mbar; range set 0000 to 1200
Set p
3-digit in mbar; range set 000 to 200
Set T auto
3-digit, one fractional part, in °C
range set 001 to 300 equals 0.1 to 30.0
Set Vapor
3-digit, one fractional part, in °C
range set 001 to 200 equals 0.1 to 20.0
Set Bath
4-digit, in 1°C increments
range set 0000 to 1800 equals 0.0 to 180.0
Set Speed
3-digit, in rpm; range set 000 to 180
Start Stop Vacuum
SV0! = OFF ; SV1! = ON
Valve
vacuum valve; VV0! = CLOSE ; VV1! = OPEN
Ventilation
aerating valve; BV0! = CLOSE ; BV1! = OPEN
Mode p const
go to p constant mode
Mode t auto
go to T auto mode
Start Stop Bath
SB0! = heater OFF ; SB1! = heater ON
Start Stop Speed
SS0! = rotation OFF ; SS1! = rotation ON
lift UP
lift DWN

FUNCTIONS AND HANDLING

5
3
4
6
2
102
1