Analogschnittstelle; Analog Interface; Interface Analogique - Ismatec BVP Process Mode D'emploi

Masquer les pouces Voir aussi pour BVP Process:

Publicité

speed IN
direction
input 1
start
+26VDC
remote
input 2
GND
8 7 6 5 4 3 2 1
15 14 13 12 11 10 9
output 2
speed OUT
output 1
+5VDC
speed intern
Digitale Eingänge (TTL-Pegel)
Digital inputs (TTL-level)
Entrées numériques (niveau TTL)
Pin 2, remote
Pin 3, start
Pin 4, direction
Pin 6, input 1
Pin 8, input 2
Pin 13, speed intern
Analoge Eingänge
Analog inputs
Entrées analogiques
Pin 5, speed IN
0–5 V
/ 0–10 V
DC
DC
0–20 mA / 4–20 mA
Universal Ausgänge (PWM)
Universal outputs (PWM)
Sorties universelles (PWM)
Pin 14, output 1
Pin 15, output 2
Analog-Ausgang
Analog output
Sortie analogique
Pin 9, speed OUT
0-10 V
/ 0-7.2 kHz
DC
36

Analogschnittstelle

Pin 1, GND (Masse)
Bezugspotential für alle anderen
Eingänge.
Pin 2, remote
Für Umschaltung zwischen ma-
nueller Bedienung und der Ana-
logschnittstelle. Zur Aktivierung
der Analog-Schnittstelle muss
Pin 2 mit Pin 1 (GND) verbunden
werden.
Pin 3, start
- Im Normalbetrieb (Pin 2 offen)
dient Pin 3 als Fußschalter
(Einstellungen siehe Seite 17)
- Im Remote-Betrieb (Pin 2 auf
GND) startet die Pumpe bei Ver-
bindung mit Pin 1 (GND)
Pin 4, direction
Wenn offen, dreht die Pumpe
im Uhrzeigersinn; wenn mit Pin
1 (GND) verbunden, dreht sie im
Gegenuhrzeigersinn
Pin 5, speed IN
Externe Drehzahlsteuerung
(0 – 5V, 0 – 10V, 0 – 20mA,
4 – 20mA)
Eingangsimpedanz und Wahl-
möglichkeiten mittels DIP-Switch
im Geräteinnern (siehe Seite 38)
Pin 6, input 1
Pin 8, input 2
Digitale Eingänge (TTL-Pegel).
Aktiv, wenn mit Pin 1 verbunden.
Können nur über RS232-Schnitt-
stelle abgefragt werden; z.B. mit
einer Software wie ProgEdit .
Pin 7, +26V
DC
Es stehen ca. +26 V
zur Verfü-
DC
gung (max. Strom 1A).

Analog interface

Pin 1, GND (ground)
Reference potential for all other
inputs
Pin 2, remote
For changing between manual
control and analog interface. For
activating the analog interface,
pin 2 must be connected with pin
1 (GND).
Pin 3, start
- In normal operation (pin 2
open) pin 3 operates the
foot-switch (settings see
page 17)
- In remote operation (pin 2 to
GND) the pump starts when
connected to pin 1 (GND)
Pin 4, direction
In the open position the pump
turns clockwise; when connected
to pin 1 (GND) it turns counter-
clockwise.
Pin 5, speed IN
External speed control
(0 – 5V, 0 – 10V, 0 – 20mA,
4 – 20mA)
Input impedance and input range
can be selected via a dip-switch
inside the pump(see page 38)
Pin 6, input 1
Pin 8, input 2
Digital inputs (TTL-level). Active
when connected to pin 1. Are
only retrievable via RS232 inter-
face; e.g. with a software like the
ProgEdit.
Pin 7, +26V
DC
About +26 V
are available (max.
DC
current 1A).

Interface analogique

Pin 1, GND (masse)
Potentiel de référence pour toutes
les autres entrées
Pin 2, remote
Pour commuter du service manuel
à l'interface analogique.
Pour activer l'interface analo-
gique, le pin 2 doit être connecté
au pin 1 (GND).
Pin 3, start
- en exploitation normale (pin 2
ouvert), le pin 3 sert d'interrupteur
au pied (selection voir en page 17)
- en exploitation à distance (pin
2 sur GND), la pompe se met
en route dès qu'elle est connec-
tée au pin 1 (GND)
Pin 4, direction
Si ouvert, le sens de rotation de
la pompe est celui des aiguilles
d'une montre; si relié avec le
pin 1 (GND), elle tourne dans le
sens contraire des aiguilles d'une
montre.
Pin 5, speed IN
Réglage externe du nombre de
tours (0 – 5V, 0 – 10V, 0 – 20mA,
4 – 20mA) Impédance d'entrée
et réglage de zone au moyen de
l'interrupteur DIP à l'intérieur de
l'appareil (v.p. 38)
Pin 6, input 1
Pin 8, input 2
Entrées numériques (niveau TTL).
Actifs si reliés au pin 1. Ne peu-
vent être interrogés que par
l'interface RS232, p.e. avec le
logiciel ProgEdit.
Pin 7, +26V
DC
Environ +26 V
sont à dispositi-
DC
on (courant maximal 1A)
MCP Process/ISMATEC SA/29.05.07/CB/GP