Note: If first name, last name, and date of birth matches an existing patient on
™
CareLink
unless you enter a different patient ID from the one in the record. In that
case, a new patient record will be created.
5
Review the patient information and tap Confirm if accurate.
Pairing the Recorder
When you have successfully created a patient record, you will need to pair the
patient's recorder with the app. The app will take you to Pair the Recorder screen
after you confirm your patient's information. Insert sensor and attach recorder per
sensor and recorder user guide.
1
Tap Search for Recorder. Bluetooth
to allow the devices to start pairing.
The app will search for available recorders within range and display their serial
numbers. If the app cannot find a recorder, the message, "No Recorders Found,"
appears. Tap More Info for instructions on how to proceed.
2
Find the serial number on the screen that matches the serial number on the
recorder's device or bag for this patient and tap that number on the screen.
3
Verify that the serial number appearing on the app screen matches the number
on the recorder's device or bag. Tap Confirm.
The mobile device and recorder will begin to pair with one another. "Pairing
device with recorder" will appear on the screen. After you have completed pairing
the recorder with the mobile device, the message, "The evaluation was
successfully started," appears on the screen.
4
Tap Done.
Home Use and General Instructions
The following section contains home use and general instructions. Some of these
instructions will appear on the app, but it is important to review these instructions with
your patient.
• Keep a log.
• You may shower and swim, up to a depth of 8ft for 30 minutes, with the device.
• Evaluation is up to 7 days.
• No MRIs, X-rays or CT scans.
• Remove the device on day 8.
• Schedule a follow up, and return the recorder and log sheet within 12 days of
starting your evaluation.
• Contact your healthcare professional or the Medtronic HelpLine with any
questions or concerns.
software, the evaluation will merge under the existing record
™*
-13-
must be turned on for the mobile device